SPECIFICATIONS:
The Delasco Electricator® is a non-sterile, reusable, electrosurgical generator, intended to be utilized for destroying small lesions and to
coagulate small bleeders by applying high frequency current through an electrode.
|
|
|
|
| Primary Power Requirements |
|
100-240V, 50/60 Hz |
|
| Maximum Supply Current |
|
1.7 amps |
|
| Fundamental Output Frequency |
|
350-490 KHz ±50 KHZ |
|
| Output Wave Shape |
|
Clipped, Damped, Sinusoidal |
|
| Maximum Output Power/Rated Load |
|
Low - 25 watts/500 ohms Bipolar - 45 watts/1000Ω (ohms) High - 45 watts/1000Ω (ohms) |
|
| Maximum Voltage Open Circuit |
|
Low - 3800Vpp High - 8500 Vpp Bipolar - 1600 Vpp |
|
| Duty Cycle |
|
30 sec ON/30 sec OFF |
|
| Display Reading |
Low |
0.1-10 range: 0.1 watt increments 10-25 range:1 watt increments |
|
High |
0-45 range: 1 watt increments |
|
Bipolar |
0-45 range: 1 watt increments |
|
| Operating Temperature Range |
|
+10 degrees to +40 degrees C |
|
| Relative Humidity |
|
10% to 95% non-condensing |
|
| Storage Restrictions |
|
-40 degrees to +70 degrees C, 95% Humidity |
|
| Protection |
|
BF Defibrillator, Class 1 |
|
| Dimensions |
|
8.75"H x 6.5"W x 3.0"D |
|
| Weight |
|
6 lbs. |
|
| High Frequency Leakage Currents |
|
Monopolar Patient Leakage (as per IEC60601-2-2,Section 19.3.101.a.1): 150 mAmps Max
Bipolar Patient Leakage (as per IEC60601-2-2,Section 19.3.101.a.3): 45mAmps Max.
|
|
| Low Frequency Leakage Currents |
|
Enclosure (as per IEC60601-1,Section 19.3, Type BF): 50µAmps Max
Monopolar Patient Leakage (as per IEC60601-1,Section 19.3,Type BF): 50µAmps Max
Bipolar Patient Leakage (as per IEC60601-1,Section 19.3,Type BF): 7.5µAmps Max
|
|
|
The Delasco Electricator® includes (below) the Electrosurgical Unit and power cord, one Handpiece, a wall mount kit
(not shown), one starter pack each of Delasco Electrolite Disposable Electrodes and Electrosurgical Handpiece Sheaths,
and an operator's manual.
|